Medical Cabinet Controller MEC100 Product Introduction
※ Cascade press control logic, used for temperature control of -70/-86 ° C cryogenic ultra-low temperature cabinets
※ Scalable recorder module, storing 100000 pieces of data, supporting PDF data export from USB drive (supporting four modes: Chinese/English/° C/° F)
※ Can directly drive 220VAC/2HP compressor, supporting 3 output configurations
※ Support MODBUS communication to output temperature data
※ Equipped with sensor self detection function, automatic alarm prompt in case of malfunction, without affecting refrigeration operation
※ Supports Celsius/Fahrenheit conversion
※ Supports one click copying, making it easy to batch modify parameters
※ Support parameter restoration, manufacturer definable, to solve related after-sales problems
Technical Parameter
※ Temperature measurement range: Cabinet temperature PT1000: -99.9 ℃~+35 ℃ Condensing, ambient temperature, high temperature evaporator NTC: -50 ℃~+110 ℃
※ Voltage Rating: 220VAC ± 10%, 50/60Hz Or 110VAC ± 10%, 50/60Hz
※ Output configuration:
High temperature compressor=17A/250VAC
Low temperature compressor=17A/250VAC
Capillary heating wire=10A/250VAC
Remote alarm=10A/250VAC
(Normally closed point output alarm)
※ Backup power supply: Input interface=1 input voltage range: 5.5-10.0VDC
※ Consumed Power of the whole machine: 5VA (Max)
※ NTC temperature measurement accuracy: ± 1 ℃, 50 ℃~70 ℃, ± 2 ℃ at -50 ℃~+50 ℃
※ PT1000 sensor temperature measurement accuracy: -90 ℃~-50 ℃, remaining ± 2 ℃
※ Operating temperature: 0 ℃~55 ℃
※ Storage temperature: -25 ℃~75 ℃